Parameter Set Mycophenolic Acid - LC-MS/MS

Parameter Set Mycophenolic Acid - LC-MS/MS

   
  • Mycophenolic acid
  • Mycophenolic acid glucuronide

Order no.: 92916/RUO, 92111/RUO, 92110/RUO

For the analysis of MPA and MPAG
6PLUS1® Multilevel Calibrator Set available
Part of the modular system of the Series A

Product Name Qty
Parameter Set - Mycophenolic Acid - LC-MS/MS
Order no.: 92916/RUO
BASIC Kit A for 200 tests - LC-MS/MS
Order no.: 92111/200/RUO
BASIC Kit A for 1000 tests - LC-MS/MS
Order no.: 92111/1000/RUO
MassTox® MasterColumn® A
Order no.: 92110/RUO

Description

The modular system MassTox® Basic Kit A in combination with the MassTox® Parameter Set for Mycophenolic Acid allows for the fast quantitative determination of mycophenolic acid and its glucuronide in serum/plasma by LC-MS/MS. The use of stable isotopically marked, co-eluating internal standards and the 6PLUS1® multilevel calibrators ensures high performance of the analysis method.

Chromatogram

Technical Data

Method of Analysis

LC-MS/MS

Parameter

Mycophenolic acid, Mycophenolic acid glucuronide

Analysis Time

1.8 min

Gradient

Starting point       60 % Mobile Phase 2
0.00→0.50 min    100 % Mobile Phase 2
0.51–1.00 min    100 % Mobile Phase 2
1.01→1.10 min     60 % Mobile Phase 2
1.11–1.81 min      60 % Mobile Phase 2

Specimen

Serum/Plasma

Sample Preparation

  • Prepare the Internal Standard Mix
  • Add 800 μl Internal Standard Mix to 12 ml Precipitation Reagent to form mixture A
  • Pipette 50 μl research sample/calibrator/MassCheck® control into a 1.5 ml reaction vial
  • Add 25 μl Extraction Buffer
  • Mix and incubate 2 min at ambient temperature (do not centrifuge)
  • Add 250 μl of mixture A and mix 30 s minimum (vortex)
  • Centrifuge 5 min at 15000 x g.
  • Dilute the supernatant with Dilution Buffer prior to injection, depending on the instrument sensitivity

Sample Stability

Research samples are stable up to 4 weeks at +2 to +8 °C. For longer storage periods keep samples frozen below -18 °C.

Injection Volume

5 - 100 μl

Ionisation

ESI positive

MS/MS-Mode

MRM