0067 0068 HPLC amiodarone desethylamiodarone plasma controls

Amiodarone, Desethylamiodarone Plasma Controls

   
  • Amiodarone
  • Desethylamiodarone

Order no.: 0067/0068

Product Name Qty
Amiodarone, Desethylamiodarone Plasma Control, Level I
Order no.: 0067
Amiodarone, Desethylamiodarone Plasma Control, Level II
Order no.: 0068

Description

Amiodarone, Desethylamiodarone Plasma Controls, lyophilised

0067 Amiodarone, Desethylamiodarone Plasma Control Level I (lyoph.), 10 x 2 ml
0068 Amiodarone, Desethylamiodarone Plasma Control Level II (lyoph.), 10 x 2 ml

These lyophilised quality controls from Chromsystems are based on human plasma and should be used according to the instruction manual. They are designed to monitor the accuracy and precision of analytical procedures in clinical diagnostics for the quantitative determination of amiodarone and desethylamiodarone in serum/plasma. While following the instructions the controls should be handled and measured in the same manner as a patient specimen.

Technical Data

Method of Analysis

HPLC

Parameter

Amiodarone, Desethylamiodarone

Concentrations

Please note: Concentration values shown here are typical. For lot specific concentration values please refer to the product insert supplied with the materials. The product inserts can also be requested from us directly or downloaded from our website after login.

SubstanceValue Level IRange Level IValue Level IIRange Level II
Amiodarone 1.00 mg/l 0.80-1.20 mg/l 2.59 mg/l 2.07-3.11 mg/l
Desethylamiodarone 1.14 mg/l 0.91-1.37 mg/l 2.85 mg/l 2.28-3.42 mg/l

Reconstitution

Carefully remove the metal seal and rubber stopper from the vials. Add exactly 2.0 ml of distilled water into each vial. Return the stopper and incubate for 10 to 15 minutes at room temperature. Swirl the vials to dissolve the contents until homogeneity.

Storage conditions

Stored at +2 to +8 °C and unopened, the lyophilised controls are stable until the expiry date specified on the vial labels. The reconstituted controls can be stored up to 2 days tightly capped at +2 to +8 C. If the controls cannot be used within this period, they should be aliquoted and stored frozen below -18 °C (up to 3 months). For more details about storage conditions please check your instruction manual.