52052-premix HPLC kit vitamin B1 vitamin B6 whole blood plasma

Vitamin B1 in Whole Blood and Vitamin B6 in Whole Blood/Plasma - HPLC

Order no.: 52052/premix
100 Tests
  • Vitamin B1
  • Vitamin B6

Each parameter safeguarded by a specific internal standard
No post-column derivatisation required
One sample preparation and one run for both parameters
Simplified sample prep with Pre-mixed Tubes

Description

Clinical relevance

Vitamin B1 (thiamine) plays an important role in deriving energy from carbohydrates and fats. The active form is thiamine pyrophosphate (TPP), a coenzyme for some enzymatic reactions in carbohydrate metabolism and for pyruvate decarboxylase. TPP is also significantly involved in the oxidative degradation of glucose. The determination of the TPP concentration is more significant than measurements of total thiamine, as by this method only the physiologically active form is taken into account.

Vitamin B6 comprises pyridoxine, pyridoxamine and pyridoxal, as well as their phosphate derivatives. It is ingested with food and transferred via several enzymatic conversions into its active form, pyridoxal-5’-phosphate (PLP). PLP is involved in nearly one hundred different enzymatic reactions, among other things, as co-factor in amino acid metabolism, and in the formation of haemoglobin or neurotransmitters in the brain.

 

Product advantages

  • High throughput due to separation in less than 3 minutes
  • Reliable and reproducible results with a specific internal standard for each parameter
  • Easy sample preparation with Pre-mixed Tubes

 

This HPLC kit facilitates the combined analysis of vitamin B1 in whole blood and vitamin B6 in whole blood or plasma. Due to the pre-mixed tubes, sample preparation consists of only a single pipetting step. The internal standard for each of the two parameters and special quality controls ensure precise and reliable quantification.

Chromsystems also offers a standard version without pre-mixed tubes.

There is also a method using UHPLC as well as automated sample preparation for HPLC and UHPLC methods available for laboratories with a higher sample throughput.

Chromatogram

Technical Data

Method of Analysis

HPLC

Number of Tests

100

Parameter

Vitamin B1, Vitamin B6

Analysis Time

< 7 min plasma; < 9 min whole blood

Limit of quantification

TPP 2.0 μg/l (whole blood)
PLP 4.5 μg/l (whole blood), 0.5 μg/l (plasma)

Linearity

PLP 500 μg/l
TPP 750 μg/l

Intraassay

CV < 3 %

Interassay

CV < 6 %

Column temperature

ambient (~25°C)

Flow rate

1.5–2.3 ml/min

Specimen

Whole Blood/Plasma

Sample Preparation

  •  In a light protected vial mix 200 μl whole blood or plasma with 100 μl Internal Standard and 300 μl Precipitation Reagent, mix for at least 30 s (vortex).
  • Centrifuge for 5 min with at least 9000 x g.
  •  Add 100 μl of Derivatisation Mix and 250 μl of supernatant to a labeled Premixed Neutralisation Tube, mix briefly.
  •  Incubate for 25 min at 60 °C (water bath).
  •  Cool sample for 10 min at +2 to +8°C, then centrifuge for 2 min with at least 9000 x g.
  • Transfer supernatant into a light protected autosampler vial, inject 25–50 μl into the HPLC system.

Sample Stability

Protect sample from light, at +2 to +8°C stable for 1 day, for longer storage deepfreeze below -18°C (maximum 2 weeks).

Injection Volume

25–50 μl

Wavelength

Start EX 320 nm, EM 415 nm,

after approx. 3.8 min switch to EX 367 nm, EM 435 nm

Please note

The freely available information on this website, in particular on the sample preparation, are not sufficient to work with our products. Please read instructions and warning notices on products and/or instruction manuals.

Kit Components

Components available separately

Calibrators

Calibrators available separately

Controls

Your Feedback is Appreciated