Parameter Set Antiepileptic Drugs All-in-One Method - LC-MS/MS

New

Parameter Set Antiepileptic Drugs All-in-One Method - LC-MS/MS

   
  • 10,11-Dihydroxycarbamazepine
  • 10-OH-carbamazepine
  • Brivaracetam
  • Carbamazepine
  • Carbamazepine-10,11-epoxide
  • Ethosuximide
  • Felbamate
  • Gabapentin
  • Lacosamide
  • Lamotrigine
  • Levetiracetam (Keppra®)
  • N-Desmethylmesuximide
  • Oxcarbamazepine
  • Perampanel
  • Phenobarbital
  • Phenylethylmalonamide (PEMA)
  • Phenytoin
  • Pregabalin
  • Primidone
  • Retigabine
  • Rufinamide
  • Stiripentol
  • Sultiame
  • Theophylline
  • Tiagabine
  • Topiramate
  • Valproic acid
  • Vigabatrin
  • Zonisamide

Order no.: 92921/XT, 92111, 92110

New method, now also including brivaracetam, perampanel, retigabine
Determination of all analytes in a single run
Now with 26 internal standards
Part of MassTox® TDM Series A

Product Name Qty
Parameter Set - Antiepileptic Drugs All-in-One Method - LC-MS/MS
Order no.: 92921/XT
BASIC Kit A for 200 tests - LC-MS/MS
Order no.: 92111/200
BASIC Kit A for 1000 tests - LC-MS/MS
Order no.: 92111/1000
MassTox® TDM MasterColumn® A
Order no.: 92110

Description

Clinical relevance

Epileptic seizures are the result of synchronous discharges of neuron groups in the brain that can lead to sudden and involuntary stereotypical behavioural or sensory disorders. Numerous types of seizures have been described, each of which requires specialised therapy. The probability of them occurring depends on a number of factors. In addition to genetic predisposition, it is primarily exogenous factors that are relevant, such as accident, thrombosis, tumours or meningitis. Therapy with anticonvulsive medication (antiepileptics) leads to a reduction in seizures and sometimes even a complete elimination of seizures in most treated patients. The precondition for the antiepileptics to have a therapeutic effect is usually patient compliance, which means regular use of the medication. Thus, monitoring of the blood levels is essential, especially during initial dose setting.

 

Product advantages

  • New method, now also including brivaracetam, perampanel, retigabine
  • Determination of all analytes in a single run
  • Now with 26 internal standards

 

With the Parameter Set Antiepileptics in serum/plasma, 29 different drugs can be measured quickly and efficiently using LC-MS/MS. Due to the careful optimisation of all kit components as well as the chromatographic separation, matrix effects (“ion suppression”) are minimised and the robustness of the method is enhanced. Sample preparation is based on a protein precipitation step. The use of stable isotopically labelled (deuterated) and co-eluting internal standards as well as 3PLUS1® multilevel calibrators ensures high precision and the reproducible and reliable quantification of the analytes. The method is comprehensively validated.

The Parameter Set is a part of the Series A modular system, which enables the analysis of more than 150 parameters without switching column or changing the mobile phases, thereby minimising required workload in the laboratory. The Basic Kit A contains all components required for sample preparation and all necessary mobile phases. The MasterColumn® A is the analytical column used for the determination of all Series A analytes. Our portfolio contains further MassTox® Parameter Sets for covering more than 150 substances and metabolites.

Chromatogram

Technical Data

Method of Analysis

LC-MS/MS

Parameter

10,11-Dihydroxycarbamazepine, 10-OH-carbamazepine, Brivaracetam, Carbamazepine, Carbamazepine-10,11-epoxide, Ethosuximide, Felbamate, Gabapentin, Lacosamide, Lamotrigine, Levetiracetam (Keppra®), N-Desmethylmesuximide, Oxcarbamazepine, Perampanel, Phenobarbital, Phenylethylmalonamide (PEMA), Phenytoin, Pregabalin, Primidone, Retigabine, Rufinamide, Stiripentol, Sultiame, Theophylline, Tiagabine, Topiramate, Valproic acid, Vigabatrin, Zonisamide

Analysis Time

4.9 min

Limit of quantification

0.004–9 mg/l

Linearity

up to 0.50–250 mg/l

Recovery

84–118 %

Intraassay

CV < 10 %

Interassay

CV < 13 %

Gradient

binary

Specimen

Serum/Plasma

Pre-analytic Treatment

Stored in the dark and cooled (+4 °C) samples are stable for at least 24 hours. For longer storage periods keep samples frozen below -18 °C.

Sample Preparation

  • Add 800 µl Internal Standard Mix to 12 ml Precipitation Reagent (= mixture A).
  • Pipette 50 µl sample/calibrator/MassCheck® control into a 1.5 ml reaction vial.
  • Add 25 µl Extraction Buffer, mix briefly (vortex) and incubate 2 min.
  • Add 250 µl of mixture A, mix 30 s (vortex) and centrifuge 5 min.
  • Dilute supernatant with Dilution Buffer (depending on instrument sensitivity) and inject into LC-MS/MS system.

Injection Volume

0.2–50 µl

Ionisation

ESI positive and negative

MS/MS-Mode

MRM

Additional Info

This method further includes 26 internal standards, which are not shown in the chromatogram for clarity.

To avoid ESI switching, the analysis of ethosuximide, phenobarbital and valproic acid can be performed separately in ESI negative mode.

Please note

The freely available information on this website, in particular on the sample preparation, are not sufficient to work with our products. Please read instructions and warning notices on products and/or instruction manuals.

Kit Components

Calibrators

Calibrators available separately

Controls

Your Feedback is Appreciated

* Required Fields