Parameter Set Antidepressants 1/EXTENDED - LC-MS/MS

Parameter Set Antidepressants 1/EXTENDED - LC-MS/MS

   
  • Citalopram
  • Desmethylfluoxetine
  • Duloxetine
  • Fluoxetine
  • Fluvoxamine
  • Mirtazapine
  • N-Desmethylcitalopram
  • N-Desmethylmirtazapine
  • N-Desmethylsertraline
  • O-Desmethylvenlafaxine
  • Paroxetine
  • Sertraline
  • Venlafaxine

Order no.: 92913/XT, 92111, 92110

Encompasses 13 analytes
3PLUS1® Multilevel Calibrator Set available
Part of the MassTox® TDM Series A

Product Name Qty
Parameter Set - Antidepressants 1/EXTENDED
Order no.: 92913/XT
BASIC Kit A for 1000 tests - LC-MS/MS
Order no.: 92111/1000
BASIC Kit A for 200 tests - LC-MS/MS
Order no.: 92111/200
MassTox® TDM MasterColumn® A
Order no.: 92110

Description

Clinical relevance

There are many drugs available for the treatment of depression. In addition to classic tricyclic antidepressant drugs, there are newer generation antidepressants: "selective serotonin re-uptake inhibitors" (SSRIs) and "selective serotonin and noradrenalin re-uptake inhibitors" (SSNRIs).

SSRIs inhibit the re-uptake of the messenger substance serotonin into the presynaptic neurons. They possess characteristics that reduce anxiety and improve mood. Usually, these positive effects start to set in about two to three weeks after starting therapy. SSNRIs are mood and motivation enhancing and act similarly to the SSRIs, but additionally block the re-uptake of noradrenalin.

Since SSRIs and SSNRIs work selectively, side effects are rare. But still, therapeutic drug monitoring is recommended, because the therapeutic endpoint is difficult to determine, and monitoring patient compliance increases the likelihood of treatment success.

 

Product advantages

With the Parameter Set Antidepressants 1/EXTENDED in serum/plasma, 13 different analytes can be measured fast and efficiently using LC-MS/MS. Due to the careful optimisation of all kit components as well as the chromatographic separation, matrix effects (“ion suppression”) are minimised and the robustness of the method is enhanced. Sample preparation is based on a protein precipitation step. The use of stable isotopically labelled (deuterated) and co-eluting internal standards as well as 3PLUS1® multilevel calibrators ensures high precision and the reproducible and reliable quantification of the analytes. The method is comprehensively validated.

The Parameter Set is a part of the Series A modular system, which enables the analysis of more than 150 parameters without switching column or changing the mobile phases, thereby minimising required workload in the laboratory. The Basic Kit A contains all components required for sample preparation and all necessary mobile phases. The MasterColumn® A is the analytical column used for the determination of all Series A analytes. Our portfolio contains further MassTox® Parameter Sets for covering more than 150 substances and metabolites.

Chromatogram

Technical Data

Method of Analysis

LC-MS/MS

Parameter

Citalopram, Desmethylfluoxetine, Duloxetine, Fluoxetine, Fluvoxamine, Mirtazapine, N-Desmethylcitalopram, N-Desmethylmirtazapine, N-Desmethylsertraline, O-Desmethylvenlafaxine, Paroxetine, Sertraline, Venlafaxine

Analysis Time

3.0 min

Limit of quantification

0.5 - 30.0 µg/l

Linearity

at least twice that of the therapeutic range

Intraassay

CV < 9 %

Interassay

CV < 9 %

Gradient

0.00-0.20 min    0 % Mobile Phase 2
0.21-0.70 min    50 % Mobile Phase 2
0.71-2.40 min    80 % Mobile Phase 2
2.41-3.00 min    0 % Mobile Phase 2

Specimen

Serum/Plasma

Sample Preparation

  • Prepare the Internal Standard Mix.

  • Add 800 μl Internal Standard Mix to 12 ml Precipitation Reagent to form mixture A.

  • Pipette 50 μl sample/calibrator/MassCheck® control into a 1.5 ml reaction vial.

  • Add 25 μl Extraction Buffer.

  • Mix and incubate 2 min at ambient temperature (do not centrifuge).

  • Add 250 μl of mixture A and mix 30 s minimum (vortex).

  • Centrifuge 5 min at 15000 x g.

  • Dilute the supernatant with Dilution Buffer prior to injection depending on the instrument sensitivity.

Sample Stability

Samples are stable for at least 24 hours when stored in the dark at +4 °C. For longer storage periods keep samples frozen below -18°C.

Injection Volume

5-40 µl

Ionisation

ESI positive

MS/MS-Mode

MRM

Additional Info

We recommend to set the scan time to a value that allows to achieve a minimum of 10 data points over the whole peak width.

Please note

The freely available information on this website, in particular on the sample preparation, are not sufficient to work with our products. Please read instructions and warning notices on products and/or instruction manuals.

Kit Components

Calibrators

Calibrators available separately

Controls

Your Feedback is Appreciated

* Required Fields