92033 LCMS TDM Series A 3PLUS1 calibrator TCA2

3PLUS1® Multilevel Plasma Calibrator Set Tricyclic Antidepressants TCA 2

Order no.: 92033
  • Clomipramine
  • Maprotiline
  • Norclomipramine
  • Normaprotiline
  • Nortrimipramine
  • Protriptyline
  • Trimipramine
Component of the Parameter Set Tricyclic Antidepressants TCA 2, available separately

Description

3PLUS1® Multilevel Plasma Calibrator Set for the Parameter Set Tricyclic Antidepressants TCA 2, lyophilised

These lyophilised calibrators from Chromsystems are based on human plasma and should be used according to the instruction manual. They are designed to calibrate the measurement method in clinical diagnostics for the quantitative determination of tricyclic antidepressants TCA 2 in serum/plasma. While following the instructions the calibrators should be handled and measured in the same manner as a patient specimen.
These calibrators are only available as multilevel calibrator set with 3 different concentration levels plus 1 blank value.

Technical Data

Method of Analysis

LC-MS/MS

Content

4 x 1 ml (lyoph.)

Parameter

Clomipramine, Maprotiline, Norclomipramine, Normaprotiline, Nortrimipramine, Protriptyline, Trimipramine

Concentrations

Please note: Concentration values shown here are typical. For lot specific concentration values please refer to the product insert supplied with the materials. The product inserts can also be requested from us directly or downloaded from our website after login.

Substance Calibrator 1
(µg/l)
Calibrator 2
(µg/l)
Calibrator 3
(µg/l)
Blank Calibrator
(µg/l)
Clomipramine 12.9 187 361 n.d.
Norclomipramine 12.8 190 365 n.d.
Maprotiline 35.4 198 365 n.d.
Normaprotiline 70.3 219 365 n.d.
Protriptyline 33.3 193 359 n.d.
Trimipramine 6.55 182 361 n.d.
Nortrimipramine 6.27 182 363 n.d.

n.d. = not detected (< LOQ)

The blank level is manufactured using the same matrix pool as used for the calibrator levels 1, 2 and 3. It was not fortified with any tricyclic antidepressants.

Reconstitution

Carefully remove the metal seal and rubber stopper from the vials. Add exactly 1.0 ml of distilled water into each vial. Return the stopper and incubate for 10 to 15 minutes at room temperature. Swirl the vials to dissolve the contents until homogeneity.

Storage conditions

Stored below -18 °C and unopened, the lyophilised calibrators are stable until the expiry date specified on the vial labels. The reconstituted calibrators can be stored in glass containers up to 1 week light protected and tightly capped at +2 to +8 C. If the calibrators cannot be used within this period, they should be aliquoted and stored frozen below -18 °C (up to 3 months). For more details about storage conditions please check your instruction manual.